Simplifying Patient-Ventilator Interactions During the Trigger Phase
- Dr. Sateesh Chandra Alavala

- Oct 28, 2025
- 1 min read
Updated: Nov 3, 2025
The following image illustrates four types of patient-ventilator interactions that can occur while triggering a breath
1) Early trigger:
A mechanical breath is initiated before the onset of neural inspiration. Revere trigger is an example of early trigger where passive mechanical inflation triggers diaphragm contraction.
2) Late trigger:
Mechanical inflation begins more than 100 milliseconds after the start of the patient's inspiration.
3) Failed Trigger:
Also known as an ineffective trigger, a patient's inspiratory effort fails to generate flow or pressure changes needed to reach the trigger sensitivity threshold, preventing the delivery of a mechanical breath.
4) False trigger (Auto trigger):
A non patient signal, such as an air leak, can cause pressure or flow changes that exceed the trigger sensitivity threshold, leading to the initiation of a mechanical breath.
Note- the following image shows pressure trigger as an example to illustrate patient-ventilator interactions





Comments